9UnpackingRemove the unit from its shipping container andensure that the following items are removed from thepackaging materials before discarding:a)
Water Connection DetailsPush-to-Connect Fitting TubingInstallationThe following instructions will apply when you need toinstall the ultrafilter and to
11Tubing Adapter Fittings1. Completely disassemble the fitting. Refer toFigure 3 to familiarize yourself with the namesof the component parts.2. Make
12INSTALLATIONCoupling InsertPush down on thumbpadto release coupling insertCoupling BodyFigure 4 Quick DisconnectNoteThe quick disconnect fittings co
INSTALLATION13CautionDo not touch the glass portion of theUV lamp. We recommend that youwear lint-free gloves when handling thelamp. The glass portion
147. Connect the UV lamp to the receptacle in theblack plastic cover. Slide the lamp completelyinto chamber and replace the black plasticcover on the
15INSTALLATIONFigure 6: Ultrafilter ConnectionsNoteThe ultrafilter comes complete withfittings installed.Tube from sanitization blockFront chassis wal
Initial Cartridge Installation (See Fig. 8)1. Open cartridge access door on the right side ofthe unit by pushing the door latch back.2. Remove a new P
Locating and Mounting UnitBench MountingPlace the EASYpure II UV/UF on a flat, level bench topthat is accessible to water, electricity and an atmo-sph
Remote Dispenser (Optional AccessoryPart No. D11981)The remote dispenser is best installed after the EASYpureII UV/UF system has been either bench or
EASYpure®II UV/UFOPERATION MANUALAND PARTS LISTSeries 1304Model D8611100-240 VoltsLT1304X1 • 4/10/03
Remote Dispenser Wall Bracket Mounting1. Determine location of bracket in relation toEASYpure II UV/UF. Using the wall bracket as atemplate, locate an
2. Install the tubing adapter into the atmospherical-ly vented drain and route the tubing to theEASYpure II UV/UF drain connector, ensuringthat there
Air Purge1. If the optional remote dispenser is connected tothe EASYpure II UV/UF, place the draw-off trig-ger into the locked position by depressing
Power SwitchThe power switch on the EASYpure II UV/UF is locatedon the left side of the unit, directly above the power cordreceptacle.Control PanelThe
DisplayIn addition to displaying the temperature compensated(25°C) resistivity in megohms-cm, the display also indi-cates operational modes and error
StandbyIn Standby (“SbY”) Mode, the pump will restart for ten min-utes out of every hour (i.e 50 minutes off, 10 minutes on),during which time the dis
25Dispensing Water from Unit1. Remove the protective cap from the filter bell.2. Place a container under the draw-off valve.3. Depress the draw-off va
SanitizationYour EASYpure II UV/UF has been shipped with a sanitiza-tion syringe (Cat. No. CMX25). When the system is new orwhenever new cartridges or
2714. Open the remote dispenser and point to drain.If no remote dispenser is present, ensure theunit draw-off valve is open and a container isplaced u
Remove Top Cover and ScrewsDaughter BoardCheck ValveMembrane Switch LeadMeter BoardRemove Front Coverand ScrewsEMI FilterDisconnect MembraneSwitch Le
Safety Information ...
4. Remove the cover by lifting it straight up. 5. Remove the 0.2 micron final filter from thedraw-off valve. Carefully remove the frontcover screws an
General Cleaning InstructionsWipe exterior surfaces with lightly dampened cloth con-taining mild soap solution. Component ReplacementCartridge Replace
Purity/Low TOC (Catalog No. D50229) cartridge,placing it in position 3. Be sure to remove andreplace one cartridge at a time.9. Close the cartridge ac
322. Depressurize the system by disconnecting unitfeedwater, opening draw-off valve and allowingwater to drain from the unit until draining ceas-es.3.
Ultrafilter Replacement1. Turn off the EASYpure II UV/UF and disconnect itfrom the power supply. Remove the power cord.2. Disconnect incoming water l
34Fuse Replacement (See Fig. 11)1. Turn off the EASYpure II UV/UF and disconnectit from the power supply. Remove the powercord.2. Pull out the fuse dr
35TroubleshootingProblem Possible Causes ActionEASYpure II UV/UF completely inactive. No electrical power to Ensure that the EASYpure II(Pump not oper
TROUBLESHOOTINGProblem (cont.) Possible Causes (cont.) Action (cont.)Reduced or no product flow 0.2 micron final filter clogged. Replace the 0.2 filte
37TROUBLESHOOTINGProblem Possible Causes ActionDisplay reads “Er1” Resistivity cell not connected to Check resistivity cell lead(Purity measurement
38ConsumablesConsumable parts are those required to support the day-to-day operation of this equipment. Barnstead International establishes two types
Dispensing Water from Remote...25Mainte
39General MaintenanceGeneral maintenance parts are defined as laboratory level repair parts which do not require great expertise orspecial tools for i
40Ordering ProceduresPlease refer to the Specification Plate for the completemodel number, serial number and series number whenrequesting service, rep
41Technical InformationWater Process Flow Diagram
42TECHNICAL INFORMATIONWiring DiagramREF. DESCRIPTION D8611 (UV/UF) C1 POWER ENTRY MODULE CEX183 E1 RESISTIVITY CELL E703X1AL1 UV LAMP LMX13PC
Barnstead International (“BARNSTEAD”) warrants that if a product manufactured by Barnstead shall be free ofdefects in materials and workmanship for on
4Safety InformationYour Barnstead EASYpure®II UV/UF has been designedwith function, reliability, and safety in mind. It is your re-sponsibility to ins
To avoid personal injury:1. Do not use in the presence of flammable orcombustible materials; fire or explosion mayresult. This device contains compone
6IntroductionProduct OverviewThe Barnstead EASYpure II UV/UF is a line-fed waterpurification system designed to be simple to use and toprovide reagent
7Dimensions and Clearance RequirementsEASYpure II UV/UF dimensions: 12 " W X 17 1/8" H X 19" D (30.5 cm x 43.5 cm x 48.3 cm).Clearance
8Environmental ConditionsOperating: 4°C - 40°C; 20% - 80% relative humidity, non-condensing. Installation Category II (overvoltage) in accordance with